Click for new scientific resources and news about Corona[COVID-19]

Paper Information

Journal:   IRANIAN JOURNAL OF MEDICAL PHYSICS   MARCH 2016 , Volume 13 , Number 1; Page(s) 36 To 42.
 
Paper: 

MEASUREMENT OF THE RADIATION DOSE RATES OF PATIENTS RECEIVING TREATMENT WITH I-131 USING TELESCOPIC RADIATION SURVEY METER

 
 
Author(s):  LAHFI YEHIA*, ANJAK OSAMA
 
* DEPARTMENT OF PROTECTION AND SAFETY, ATOMIC ENERGY COMMISSION OF SYRIA, DAMASCUS, SYRIA
 
Abstract: 

Introduction: In order to discharge the patients receiving treatment with large radiation doses of 131I for thyroid cancer, it is necessary to measure and evaluate the external dose rates of these patients. The aim of the study was to assess a new method of external dose rate measurement, and to analyze the obtained results as a function of time.
Materials and Methods: In this study, a telescopic radiation survey meter was utilized to measure the external dose rates of a sample population of 192 patients receiving treatment with high-dose 131I at one, 24, and 48 hours after dose administration.
Results: The proposed technique could reduce the occupational radiation exposure of the physicist by a factor of 1.16. Moreover, the external dose rates of both genders rapidly decreased with time according to bi-exponential equations, which could be attributed to the additional factors associated with iodine excretion, as well as the physiology of the body in terms of 131I uptake.
Conclusion: According to the results of this study, telescopic radiation survey meter could be used to measure the external dose rates of patients receiving treatment with 131I. Furthermore, the average difference in the radiation exposure between female and male patients was calculated to be less than 17%.

 
Keyword(s): RADIATION PROTECTION, RADIATION DOSE, IODINE, RADIATION MONITORING
 
 
References: 
  • ندارد
  •  
 
Click to Cite.
APA: Copy

LAHFI, Y., & ANJAK, O. (2016). MEASUREMENT OF THE RADIATION DOSE RATES OF PATIENTS RECEIVING TREATMENT WITH I-131 USING TELESCOPIC RADIATION SURVEY METER. IRANIAN JOURNAL OF MEDICAL PHYSICS, 13(1), 36-42. https://www.sid.ir/en/journal/ViewPaper.aspx?id=568966



Vancouver: Copy

LAHFI YEHIA, ANJAK OSAMA. MEASUREMENT OF THE RADIATION DOSE RATES OF PATIENTS RECEIVING TREATMENT WITH I-131 USING TELESCOPIC RADIATION SURVEY METER. IRANIAN JOURNAL OF MEDICAL PHYSICS. 2016 [cited 2021May12];13(1):36-42. Available from: https://www.sid.ir/en/journal/ViewPaper.aspx?id=568966



IEEE: Copy

LAHFI, Y., ANJAK, O., 2016. MEASUREMENT OF THE RADIATION DOSE RATES OF PATIENTS RECEIVING TREATMENT WITH I-131 USING TELESCOPIC RADIATION SURVEY METER. IRANIAN JOURNAL OF MEDICAL PHYSICS, [online] 13(1), pp.36-42. Available: https://www.sid.ir/en/journal/ViewPaper.aspx?id=568966.



 