Journal:   JOURNAL OF OPHTHALMIC AND VISION RESEARCH   2015 , Volume 10 , Number 1; Page(s) 43 To 48.
 
Paper: 

INTRAOCULAR PRESSURE MEASUREMENT BY THREE DIFFERENT TONOMETERS IN PRIMARY CONGENITAL GLAUCOMA

 
DOI: 

10.4103/2008-322X.156105

 
Author(s):  ZAREEI ATHAR*, RAZEGHINEJAD MOHAMMAD REZA, NOWROOZZADEH MOHAMMAD HOSEIN, MEHRABI YADOLLAH, AGHAZADEH AMIRI MOHAMMAD
 
* POOSTCHI EYE RESEARCH CENTER, ZAND STREET, SHIRAZ 71349, IRAN
 
Abstract: 

Purpose: To determine the agreement between intraocular pressure (IOP) measurements using an automated non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), and the ocular response analyzer (ORA) in subjects with primary congenital glaucoma (PCG).
Methods: Twenty-nine eyes of 17 PCG patients underwent IOP measurements using NCT, GAT and ORA.
Variables obtained by the ORA were corneal?compensated IOP (IOPcc), Goldmann?correlated IOP (IOPg), corneal hysteresis (CH), and corneal resistance factor (CRF). A difference more than 1.5 mmHg for IOP was considered as clinically relevant.
Results: Mean age of the patients was 12 years. Mean IOP (±standard deviation, SD) was 15.3±2.8 mmHg (GAT), 15.5±6.0 (NCT), 19.2±7.0 (IOPg), and 21.1±7.9 (IOPcc); (P=0.001). Except for NCT vs. GAT (P=1.0), the average IOP difference between each pair of measurements was clinically relevant. The 95% limits of agreements were -10.2 to 10.3 mmHg (NCT vs. GAT), -7.8 to 15.3 (IOPg vs.
GAT), and -8.1 to 19.0 (IOPcc vs. GAT). The differences in IOP measurements increased significantly with higher average IOP values (r=0.715, P=0.001, for NCT vs. GAT; r=0.802, P<0.001, for IOPg vs. GAT; and r=0.806, P<0.001, for IOPcc vs. GAT). CH showed a significant association with differences in IOP measurements only for IOPcc vs. GAT (r=0.830, P<0.001).
Conclusion: Mean IOP obtained by NCT was not significantly different from that of GAT, but ORA measured IOPs were significantly higher than both other devices.

 
Keyword(s): GOLDMANN APPLANATION TONOMETER, INTRAOCULAR PRESSURE, NONCONTACT TONOMETER, OCULAR RESPONSE ANALYZER, PRIMARY CONGENITAL GLAUCOMA
 